Warning W05

Warning W05

      a) 此故障有可能是操作不当引起,首先判断故障原因为“脚踏未连接”,请确认“脚踏开关插头”已被正确插入主机中。
注意:
① 插入 “脚踏开关插头”时,不要使用蛮力,注意和机壳对应插座红点对齐后,轻轻推入即可。“脚踏开关插头”支持热插拔。
② “脚踏开关”为随机附件。
     b) 将主机脚踏插座1、3短路,如果此时该报警消除,则判断脚踏插头故障,检查脚踏插头。
 
     c) 如果此报警未消除,则将仪器后壳拆开后,使用镊子短路主机主控板P6.4、P6.6。如果此报警仍未消除,则判断主机主控板故障,考虑更换主机主控板。
     d) 如果上述步骤仍不能解决问题,则考虑为仪器线路故障。应检查“主机主控板D43”状态,该指示灯在“脚踏开关插头正确插入”时为“熄灭”状态。如果指示灯状态与实际情况不符,则应按照“接线图”查   找对应线束是否松动,是否有断线故障。(“脚踏开关插头正确插入”情况下P6.4、P6.6短路,可以用万用表辅助测试线路通断)
     a) This failure may be resulted from the improper operation. First, judge whether the failure is ascribed to “foot switch is not connected”. Please confirm whether the “foot switch plug” has been correctly inserted into the main console.

Note:

① Do not apply too much force while inserting the “food switch plug”. After aligning to the corresponding socket red point of shell, push in the plug gently. “Foot switch plug” supports the hot plug.

②  “Foot switch plug” is an accompanying accessory.

     b) Short circuit the main console food switch sockets 1 and 3. If the alarm is eliminated at this time, judge it is the failure of food switch plug and check the food switch plug.


      c) If the alarm is not eliminated, take apart the back shell of instrument and then short circuit P6.4 and P6.6 with tweezers; If, however, the alarm is not eliminated, it should be the failure of main control board of main console, and try to replace the main control board of main console.

     d) The failure may be ascribed to the instrument line if the alarm is not eliminated after the above steps. Check the state of “main control board D43 of main console” (the indicator lamp is in the “off”  after the “foot switch plug is correctly inserted”). If the state of indicator lamp does not comply with the reality, confirm whether the corresponding wiring harness is loose and whether any line is broken according to the “wiring diagram”. (Where P6.4 and P6.6 are short circuited after “foot switch plug is correctly inserted”, use a multimeter to help test line connection/disconnection.)




    • Related Articles

    • Warning E05: the failure of water pump

      Prepare tools as below:    Multimeter,tweezers The inspection procedure of warning E05 When the machine prompt E05,first we need to enter the interface of "the engineering data checking & disable warning pop-up message",check the data of “water ...
    • Warning W02

      a) 此故障有可能是操作不当引起,首先判断故障原因为“水箱水位不足”,请使用加水工具对产品加入适量的去离子水。可以通过产品侧面液位观察窗确认当前水位状态。 注意: ① 必须加入去离子水,不能加入矿物质水或者自来水。 ② 允许带电加水,当水位达到要求,提示框自动消除。 ③ 加入过量去离子水有可能造成溢流,请注意。 b) 如果上述步骤不能解决,则将仪器后壳拆开,用镊子短路主机主控板J4.3、J4.4,此时该报警应消除。如果报警未消除,则主机主控板故障,考虑更换主机主控板。 ...
    • Warning W04

      a) 此故障有可能是操作不当引起,首先判断故障原因为“治疗室门已打开”。 1. 将主机门联锁插座1、2短路,3、4短路,此时该报警消除,则判断门联锁插头故障,按照步骤2、3进行检验,否则进行步骤b)。        2. 如果您未对门联锁插头进行改装,请确认使用了随机门联锁插头;默认为治疗室门关闭状态。       ...
    • Warning W01

      ​​ a) 此故障有可能是操作不当引起,首先判断故障原因为“门联锁插头未插入”,请确认“门联锁插头”已被正确插入主机中。 注意: ① 插入 “门联锁插头”时,不要使用蛮力,注意和机壳对应插座红点对齐后,轻轻推入即可。“门联锁插头”支持热插拔。 ② “门联锁插头”为随机附件。   b) 如果上述步骤不能解决问题,则准备镊子(或其他短路端子),短路1、2端子,如果报警消除,则判断为“门联锁插头”故障。   ...
    • Warning W03

      a) 此故障有可能是操作不当引起,首先判断故障原因为“急停按钮被按下”,按照急停按钮上标注箭头方向旋转急停按钮,则急停按钮被弹起,提示信息消除。 b) 如果上述步骤仍不能解决问题,则将仪器后壳拆开后,用镊子短路P8.1、P8.2,此时该报警应消除。如果报警未消除,则主机主控板故障,考虑更换主机主控板。 c) 如果报警消除,则考虑为仪器线路故障。应检查“主机主控板D40”状态,该指示灯在“急停被按下”时为“点亮”状态。如果指示灯状态与实际情况不符,则应按照“接线图”查找 ...