Warning E18: output pulse width is too large

Warning E18: output pulse width is too large

a) This failure means the “too large of laser emission pulse width”. Use it again after restarting the instrument. If this failure occurs again, find out the problem. If no problem is discovered, it will be caused by interference. At this time, continue to use it by ignoring this problem.

b) Plug off the main control board J12 and restart the instrument for retest. If this failure is eliminated, try replacing the DB9 wiring harness and laser power supply.

c) If this failure is not eliminated, try replacing the main control board of main console.
    • Related Articles

    • Warning E16: detection to laser emission current is too large/small

      a) This failure means “current testing – laser emission current is tested too large/small”. Use it again after restarting the instrument. If this failure still occurs, find out the problem. At this time, continue the use by ignoring this problem. ...
    • lower energy output of handpiece

      1. For handpiece having been delivered but having not been modified a) Shut down the machine and observe through the treatment head window whether the semiconductor laser unit has black point or any other burning trace. If any, replace the ...
    • No laser output

      Tips: Scope of application: Engineer Applicable grade: Professional Service tools: Multimeter  Apply text preconditions: 1、Renuva® works normal and there are no other warning; 2、The scanning galvanometer works normally; Step1. Checking the laser fan ...
    • Warning E16 电流检测——出光电流检测过大/过小

      a) 此故障代表“电流检测——出光电流检测过大/过小”。请重启仪器后再次使用,如果此故障仍出现,则查找问题,否则为干扰导致,可忽略继续使用。 b) 首先断电后,目视手柄治疗窗口,观察半导体激光器是否有烧毁损伤痕迹(单巴有明显黑点或黑条),如果有,则为半导体激光器损坏,需要更换半导体激光器。 c) 可以用相纸测试,把能量密度调至5-10J/cm²,试验是否出光。 d) 出光情况下,请测试200ms脉宽,1Hz,出光1次,10J/cm²和30 ...
    • Warning W08