prompt the handpiece keeping in/out of the handpiece holder all the time on the screen

prompt the handpiece keeping in/out of the handpiece holder all the time on the screen

a) Test whether the handpiece is inserted with the photoelectric sensor in handpiece holder. First, confirm no other sheltering object or luminous body is near to the photoelectric sensor.
b) After putting headpiece in handpiece holder, check whether the D46 indicator lamp of main control board changes from on to off. If there is no change, check whether wire harness of sensor has poor contact or falls and try to replace the photoelectric sensor. Otherwise, it should be the damage of circuit board. Replace the circuit board at this time.
    • Related Articles

    • The handpiece display screen is abnormal

      a)Where there is a blank screen and handpiece indicator lamp is not on, try plugging off and then plugging in the FFC connecting line again.   b)Where there is a blank screen and handpiece indicator lamp is on, try replacing the handpiece display ...
    • Warning E08: handpiece cooling failure and component emittng heat

      Prepare tools as below: Multimeter,hair dryer   Under the condition of continuous and normal laser shot from the handpiece for 1 minute,if the “Temp” on screen remains more than 10℃ for a long time,then the device prompts E08. 1. We touch the head of ...
    • Abnormal display on screen of main console (black screen)

      a) Remove the display maintenance window above the machine. After startup, observe whether there is the flickering indicator lamp on the Android LCD screen communication board. If no indicator lamp flicker, replace the Android LCD screen ...
    • lower energy output of handpiece

      1. For handpiece having been delivered but having not been modified a) Shut down the machine and observe through the treatment head window whether the semiconductor laser unit has black point or any other burning trace. If any, replace the ...
    • TEC in handpiece can't refrigerate

      a) First, confirm whether there is a problem in temperature test. The temperature of 4-7℃ is not displayed in interface. b) Under the ready state, test the voltage at two ends of main control board J6 of main console. It should be ≥5V and the ...