Warning W13: maintenance window is opened

Warning W13: maintenance window is opened

a) This failure may be resulted from the improper operation. First, judge whether the failure is ascribed to “maintenance window is opened”. Try closing the maintenance window.

b) If, however, this alarm still cannot be removed, try pressing the detector switch to eliminate it. There are two detector switches, respectively on the front cover plate and back cover. The alarm will be eliminated after pressing the two detector switches.
    • Related Articles

    • Warning W04: treatment room door is opened

      a) This failure may be resulted from the improper operation. First, judge whether the failure is ascribed to “treatment room door is opened”.         1. Short circuit main console gate interlock plug sockets 1 and 2 and main console gate interlock ...
    • Warning W08

      a) 此故障代表“水箱温度超出上下限”,极有可能是手柄长时间以高能量出光导致的。此提示会在水箱温度恢复正常后自动消失。 b) 如果该故障提示很容易出现(持续工作1h以内就会出现),则考虑为仪器故障。 c) 首先查看水质检测板显示水温,如果未出现温度跳变现象,则排除温度传感器接触不良问题,否则需检查温度传感器或更换水质检测板。 ...
    • Warning W04

      a) 此故障有可能是操作不当引起,首先判断故障原因为“治疗室门已打开”。 1. 将主机门联锁插座1、2短路,3、4短路,此时该报警消除,则判断门联锁插头故障,按照步骤2、3进行检验,否则进行步骤b)。        2. 如果您未对门联锁插头进行改装,请确认使用了随机门联锁插头;默认为治疗室门关闭状态。       ...
    • Warning E11

      说明: 由于版本不兼容性,主机软件版本L0-A-1.0.129、L0-A-1.0.133、L0-A-1.0.134不兼容手柄软件版本L4-A-1.0.001(或基于此版本定制版本) 。维修前请核对是否为使用此版本搭配导致。 a) 此故障代表“与手柄通讯异常”。 b) 可以通过尝试更换手柄来缩小诊断范围。 c) 如果判断为主机问题,且检查手柄显示屏或指示灯有显示,则检查主机主控板P5.7、P5.8与主机端手柄插口,是否断线或接触不良。 ...
    • Warning W02

      a) 此故障有可能是操作不当引起,首先判断故障原因为“水箱水位不足”,请使用加水工具对产品加入适量的去离子水。可以通过产品侧面液位观察窗确认当前水位状态。 注意: ① 必须加入去离子水,不能加入矿物质水或者自来水。 ② 允许带电加水,当水位达到要求,提示框自动消除。 ③ 加入过量去离子水有可能造成溢流,请注意。 b) 如果上述步骤不能解决,则将仪器后壳拆开,用镊子短路主机主控板J4.3、J4.4,此时该报警应消除。如果报警未消除,则主机主控板故障,考虑更换主机主控板。 ...