Warning W11: the tested water quaity is beyond the standard
a) This failure may be resulted from the improper operation. First, judge whether the failure is ascribed to “the water quality is tested out of standard”, and then replace the cooling water.
b) The instrument is designed with water quality pre-alarm. When the
flickering sign occurs at the right upper corner of operating interface, replace the cooling water timely.
c) If the filter element turns yellow, replace the filter element timely; or when the black particle occurs, check whether water quality reaches the standard (TDS value of new water is generally ≤20).
d) If the problem cannot be resolved after the above steps, it should be the instrument line failure. Try replacing the TDS probe or water quality testing board.
Related Articles
Warning W02: water level insufficiency of water tank
a) This failure may be resulted from the improper operation. First, judge whether the failure is ascribed to the “water level insufficiency of water tank”. Add the moderate amount of deionized water to the Product with the corresponding tool. Confirm ...
Warning E05: the failure of water pump
Prepare tools as below: Multimeter,tweezers The inspection procedure of warning E05 When the machine prompt E05,first we need to enter the interface of "the engineering data checking & disable warning pop-up message",check the data of “water ...
Warning E04: water velocity is lower than the lower limit
a) This failure means that “water velocity is lower than the lower limit”. It occurs easily after water replacement of instrument and could be resolved by restarting the instrument for several times. b) Replace filter element if it is not caused by ...
Warning W08: temperature of water tank exceeds the scope
a) This failure means that “temperature of water tank exceeds the scope”. It is very likely to be caused by the long-time high-energy laser emission. The prompt will disappear automatically after water tank temperature is recovered. b) If the failure ...
Water way alarm during self-test
Fault phenomenon: Self-test alarm when power-on, Jump out icon "Water velocity is abnormal" Make judgments according to the alarm on the self-check interface: State 1: There is an alarm in Water Flow Testing No alarm reported in Water Pump Testing ...